5091-1596E.pdf

(471 KB) Pobierz
Agilent
High Accuracy and
Fast RF Inductor Testing
Application Note 369-10
Agilent 4285A
Precision LCR Meters
935955239.026.png 935955239.027.png 935955239.028.png 935955239.029.png 935955239.001.png 935955239.002.png 935955239.003.png 935955239.004.png 935955239.005.png 935955239.006.png 935955239.007.png 935955239.008.png 935955239.009.png 935955239.010.png 935955239.011.png 935955239.012.png 935955239.013.png 935955239.014.png 935955239.015.png 935955239.016.png
 
Introduction
Problems and the solutions offered
by the Agilent 4285A
RF inductors are employed in electronic equipment such
as VCRs and intermediate frequency (IF) circuits for
AM/FM radios, mobile radios, pagers and code-less
telephone, which require RF inductors to be smaller,
lower cost, and have higher reliability. This application
note describes solutions offered by the Agilent 4285A
precision LCR meter for realizing these requirements.
Information for accurate and fast RF inductor testing,
and for practical simple test systems are discussed.
1. Measurement accuracy and measurement speed
Stable inductance (L) and quality factor (Q) low
inductance measurements are difficult to make, because
measuring low inductance values means measuring low
impedance, values. If a higher measurement frequency
(MHz) is issued to increase the measured impedance,
the LCR meter’s measurement accuracy is worse,
so accurate measurement data cannot be obtained.
The 4285A can perform reliable and fast measurements
with 0.5% inductance measurement accuracy even in the
MHz frequency range, and with a 30 ms measurement
speed. Formerly LCR meters could not measure
accurately enough for testing high Q inductors, and
Q meters required long measurement time and high
operator skill. Table 1 lists the 4285A’s key specifications.
2. Measurement condition setup
There are many types of RF inductors, so the measurement
frequency range for an inductor is selected from
across a frequency range spanning kHz through MHz,
depending on the inductance value to be measured and
the application in which it will be used. The 4285A
covers a measurement frequency range of from 75 kHz
to 30 MHz, including frequencies defined in the MIL
C-15305D standard (79.6 kHz, 252 kHz, 796 kHz, ..25.5 MHz)
A constant current test signal gives the best inductance
measurement results, because inductive devices are
current mode devices and L and Q values depend on
current levels rather than voltage levels. Q meters do
not offer current test signals, and former LCR meters
used voltage test signals, not current test signals.
The 4285A offers both voltage and current test signal
level selections. Furthermore, its auto level control (ALC)
function allows constant test signal level measurements.
Table 1. 4285A key specifications
Agilent 4285A (standard)
Measurement method
Auto balancing bridge
Measurement frequency
75 kHz to 30 MHz with 100 Hz resolution
Basic accuracy (typical)
10 mH
L: ±0.18%
@100 kHz
Q: ±5% @Q=30
1 µH
L: ±0.5%
@10 MHz
Q: ±10% @Q=30
Measurement speed (typical)
30 ms/65 ms/ 200 ms
Test signal level
V: 5 mVrms ~ 2 Vrms
(10 mVrms ~ 1 Vrms in constant mode)
I: 200 µ/Arms ~ 20 mArms
(100 µArms ~ 20 mArms in constant mode)
V/I monitors are available
2
935955239.017.png 935955239.018.png 935955239.019.png 935955239.020.png 935955239.021.png
 
Example of an accurate and fast
measurement system
3. Functions for a test system
Formerly an LCR meters’ cable extension capability was
limited to an upper measurement frequency limit of
approximately 15 MHz, and they could not fully compensate
for errors due to extension cables and test fixtures. So,
the LCR meter performance was limited when used in a
measurement test system. The 4285A can use 1 m or 2 m
test leads over its full frequency range, and its open/short
/load correction function can compensate for measurement
errors due to extension leads. You can obtain accurate
measurement values by performing error correction at
the actual measurement frequencies (for example the
frequencies defined in MIL standards).
This paragraph describes an example of L-Q measurement
and sorting system for RF inductor production line
applications, and gives some techniques for designing
such systems.
Figure 1 shows a simplified model of an RF inductor
test system. In this system, the 4285A measures L and Q
and performs BIN sorting using its internal comparator.
Equipment used in this system is as follows (refer to
‘Appendix A. Ordering Information’ for details).
• 4285A precision LCR meter with option 201 handler
interface
• 16048A test leads (l m, BNC connectors )
• Handler interface cable (Amphenol 36 pin connectors
(male))
• Automatic component handler (compatible with
4285A option 201)
An internal comparator and an optional automatic
handler interface are available for setting up RF inductor
test systems.
The following are technical hints for designing a test
system, and for maximizing the 4285A’s performance.
1. Cable connections
Use Agilent test leads to extend the measurement
terminals of the 4285A to the handler’s contact
electrodes.
You can select 1 m or 2 m test leads, however, 1 m test
leads are recommended to minimize the error caused
by the addition of extension cables.
2. Extension leads and contact electrodes
• The 16048A test leads will extend the signal path
from the 4285A measurement terminals to the end
of cables while maintaining a Four-Terminal Pair
configuration.
Figure 1 . Example of an accurate and
fast L-Q measurement system
Use connector plate (Agilent PN 16032-60001,
furnished with the 16048A) to change the terminal
configuration and extend the leads to the contact
electrodes as shown in figure 2.
Figure 2 . Contact electrodes
3
935955239.022.png
Extension leads and contact electrodes which are not a
Four-Terminal Pair configuration cause measurement
errors, because of residual impedance and noise. Thus,
pay close attention to the following items. Especially
when performing low inductance measurements, the
following must be closely considered, because for low
impedance measurements, measurement repeatability
depends on contact electrode stability rather than the
4285A’s inductance measurement repeatability.
• To shield against noise, the outer conductors of the
extension leads must be connected to each other at
a point as near as possible to the test device.
• To reduce residual impedance, the extension leads
must be as short as possible.
As shown in figure 4, leave the contact electrodes open,
then perform an open correction measurement to
compensate for the stray admittance of extension leads
and the contact electrodes.
The short correction function compensates for the
residual impedance of extension leads and contact
electrodes, and is very important for L and Q
measurements. Short the electrodes to each other as
directly as possible, then perform a short correction
measurement. If you want to base your measurement on
a working standard as a reference, then we recommend
performing a load correction measurement in addition
to an open/short correction measurement. The load
correction function compensates for measurement errors
of a test device by using load correction coefficients
obtained from the measured values and the reference
values of a working standard (load). Input the reference
values of the working standard such as; L and Q values
(C and D in figure 3). Connect the working standard
to the contact electrodes, then perform the load
correction measurement.
For stable residual impedance, the layout, positioning,
and length of the extension leads and contact electrodes
must not be changed, and careful consideration must
be given to eliminating mechanical vibration.
3.Error correction
The Agilent 4285A’s powerful error correction function
can be used to compensate for the residual impedance
of extension leads and contact electrodes to give more
accurate impedance (L and Q) measurement values.
When using the 4285A, the reference values of the
working standard can be obtained at the actual test
frequency and signal level. Measure the working standard
using the 4285A with a test fixture (such as the, 16047E
or 16034G) setup under actual measurement conditions,
and input these values as the reference values.
First, set the cable length function to 1 m to correct
for the error caused by 1 m test leads (A in figure 3).
Open/short/load correction should be performed
at the actual measurement frequencies. Up to seven
correction frequencies can be set (B in figure 3), so
the six frequency points specified in MIL C-15305D
can be covered.
Figure 3. Correction function setup
Figure 4. Performing correction measurements
4
935955239.023.png 935955239.024.png
4. Measurement conditions and comparator setup
Figure 6 shows a sample timing chart for an interface.
The total testing time for a device is as follows.
The measurement conditions, including frequency and
signal level, should be set to appropriate values.
Measurement Time (30 ms @SHORT INTEG TIME) -
Calculation/Comparison time + Device change/Sort Time.
If necessary, a delay time (l ms - 60 s) can be inserted
before the measurement time (Tm1 in figure 6) to allow
for a mechanical contact check within the handler.
Figure 5 shows a comparator setup to sort test devices
into BIN 1 through BIN 9 by L value and into IN/OUT
by Q value.
6. Memory card
5. Interface timing design
Setup data including measurement conditions and
comparator limit values can be stored in a memory
card. Using this function, all setup data can be recalled
with just a few key strokes, this is especially useful on the
production line.
The 4285A option 201 handler interface has the capability
for adjusting the control signal timing between the 4285A
and a handler. The interface is triggered by an
(/EXT_TRIG) signal from a handler, and outputs an end
of analog measurement signal (/INDEX), an end of
digital computation and comparison signal (/EOM),
and BIN sort data (/BIN 1 through/BIN 9). Each output
and input is isolated using an optocoupler, and the
outputs are open collector.
Figure 5 : Comparator setup
Figure 6 : Sample interface timing chart
5
935955239.025.png
Zgłoś jeśli naruszono regulamin