5091-4480E.pdf

(115 KB) Pobierz
Agilent AN 346-2
Balanced Circuit Measurement with an
Impedance Analyzer/LCR Meter/Network
Analyzer
Application Note
Introduction
How a balanced circuit differs from an
unbalanced circuit
A balanced circuit has its electrical
midpoint grounded. An unbalanced
circuit, however, has one side ground-
ed. A balanced circuit is typically
used in communications equipment
because a balanced circuit has the
advantage of better spurious noise
suppression.
Measuring a balanced circuit with an
unbalanced measurement instrument
A balanced circuit cannot be directly
measured with an unbalanced meas-
urement instrument because of the
difference in their configuration.
Selecting a Balun
There are several types and brands of
balun transformers. When selecting a
balun, ensure that frequency is com-
patible with your measurement
requirements. When you measure the
impedance parameters of a balanced
circuit, you don’t have to use the
balun which has the same impedance
with the circuit under test. However,
when you measure the transmission
or reflection of it, you have to use a
balun which has the same impedance
with the circuit under test to keep
impedance matching. Table 1 shows
recommended balun transformers.
When measuring balanced circuits,
the unbalanced measuring instrument
requires a balun (balanced to unbal-
anced) transformer. A balun is a type
of impedance-matching RF trans-
former.
Figure 1 shows a balanced cable
which is an example of a balanced
circuit. The voltages of the cable’s two
conductors are at every point equal
in amplitude and opposite in phase.
Figure 2 shows an unbalanced cable
which is an example of an unbal-
anced circuit. Most measurement cir-
cuits in Agilent Technologies’ imped-
ance analyzers and LCR meters are
unbalanced.
Figure 3 shows the configuration for
measuring a balanced circuit with an
unbalanced instrument.
Note: In balanced cable measure-
ments, residual current in the balun
or the measuring instrument can
cause measurement errors. To reduce
the degree of error, perform
open/short and load compensation at
the measurement terminals of the
balun.
+V
-V
+V
+V
High
terminal
V
-V
0
Low
terminal
A
Figure 1. Balanced cable
Balanced cable
Measurement Instrument
Figure 3. Balanced cable measurement configuration
+V
0V
Figure 2. Unbalanced cable
935955129.105.png 935955129.116.png 935955129.126.png 935955129.137.png 935955129.001.png 935955129.012.png 935955129.023.png 935955129.034.png 935955129.045.png 935955129.056.png 935955129.062.png 935955129.063.png 935955129.064.png 935955129.065.png 935955129.066.png 935955129.067.png 935955129.068.png 935955129.069.png 935955129.070.png 935955129.071.png 935955129.072.png 935955129.073.png 935955129.074.png 935955129.075.png 935955129.076.png 935955129.077.png
 
Measurement Configuration with a
Balun and Compensation
Impedance measurement configuration
with Agilent 4294A impedance analyzer
Figure 4 shows impedance measure-
ment configuration (1)/(2) with the
4294A.
Short standard
PN 1250-0929
Transmission measurement configuration
with a network analyzer
Figure 7 shows transmission meas-
urement configuration (3) with a net-
work analyzer.
50 Ω load standard
PN 11593A
3. Remove the connector and connect
the DUT. Measure the DUT.
Impedance measurement configuration
with the Agilent 4395A(#010)
Figure 6 shows impedance measure-
ment configuration (2) with the
4395A.
To calibrate/compensate:
Short the terminals closest to the
DUT to the signal out and to the test
port, then perform response/thru
calibration.
To calibrate/compensate for (1):
1. Perform open, short, and load
compensation at the balanced termi-
nals of the 16314-60011. Use the
furnished compensation standards of
the 16314-60011.
To calibrate/compensate:
Refer to “To calibrate/compensate for
(2)” of impedance measurement con-
figuration with 4294A impedance
analyzer.
Standards
0 Ω
PN 04191-85300
0S
PN 04191-85302
50 Ω
PN 04191-85301
Table 1. Recommended balun transformers
To calibrate/compensate for (2):
1. Assemble a female BNC connector
as shown in Figure 5.
Unb/Bal. (W)
Bandwidth
Type No.
Suppliers
50:50
100 Hz to 10 MHz
16315-60011
Agilent Technologies
50:100
100 Hz to 10 MHz
16316A
Agilent Technologies
2. Perform open, short, and load cali-
bration at the BNC connector using the
following BNC Calibration standards:
50:600
100 Hz to 3 MHz
16317A
Agilent Technologies
50:50
0.1–125 MHz
0001BB
North Hills Signal Processing
50:75
0.1–125 MHz
0101BB
North Hills Signal Processing
50:100
0.1–125 MHz
0300BB
North Hills Signal Processing
50:600
0.1–65 MHz
0700BB
North Hills Signal Processing
Agilent 4294A
75:50
0.1–100 MHz
1000BB
North Hills Signal Processing
75:75
0.1–100 MHz
1100BB
North Hills Signal Processing
75:100
0.1–100 MHz
1300BB
North Hills Signal Processing
75:600
0.1–60 MHz
1700BB
North Hills Signal Processing
Lc
Lp
Hp
Hc
Configuration (2)
Convert to 7-mm terminal
DUT
BNC(f)
Configuration (1)
APC7-N(f)
Adapter
P/N 11524A
N(m)-BNC(m)
Adapter
Shield connected
to instrument ground
Agilent 16315-60011,
16316/7A
Agilent 42942A
terminal Adapter
Connect wire and solder it to Balun
DUT
7-mm terminal
Wire
North Hills
Balun
N(m)-BNC(f)
Adapter
P/N 1250-0780
BNC(m)-Dual Banana
Jack Adapter
Shield connected
to instrument ground
DUT
UNBAL
50
Shield connected
to instrument ground
Unbalanced
16314-60011
Balanced
50
BAL
GROUND
14-mm
Figure 4. Measurement configurations with 4294A
2
935955129.078.png 935955129.079.png 935955129.080.png 935955129.081.png 935955129.082.png 935955129.083.png 935955129.084.png 935955129.085.png 935955129.086.png 935955129.087.png 935955129.088.png 935955129.089.png 935955129.090.png 935955129.091.png 935955129.092.png 935955129.093.png 935955129.094.png 935955129.095.png 935955129.096.png 935955129.097.png 935955129.098.png 935955129.099.png 935955129.100.png 935955129.101.png 935955129.102.png 935955129.103.png 935955129.104.png 935955129.106.png 935955129.107.png 935955129.108.png 935955129.109.png 935955129.110.png 935955129.111.png 935955129.112.png 935955129.113.png 935955129.114.png 935955129.115.png 935955129.117.png 935955129.118.png
 
Wires
Soldering
Balun
BNC(m)-BNC(f)
Binding Post Terminal
Agilent 4395A-010
Figure 5. Assembling BNC connector
DUT
BNC(f)
7-mm N(f)
adapter
P/N 11524A
N(m)-BNC(m)
adapter
Shield connected
to instrument ground
Agilent 16315-60011,
16316/7A
Agilent 43961A
impedance test kit
Connect wire and solder it to balun
DUT
7-mm terminal
Wire
Shield connected
to instrument ground
N(m)-BNC(f)
adapter
P/N 1250-0780
BNC(m)- dual banana
jack adapter
North Hills
balun
Figure 6. Measurement configuration (2)
Instrument
Signal
out
Test
port
N or BNC
N or BNC
DUT
N(m)-BNC(m)
P/N 1250-0780
N(m)-BNC(m)
P/N 1250-0780
Agilent 16315-60011,
16316/7A
Agilent 16315-60011,
16316/7A
N(m)-BNC(m)
P/N 1250-0780
or
N(m)-BNC(m)
P/N 1250-0780
or
BNC(m) - dual banana
jack adapter
BNC(m) - dual banana
jack adapter
BNC(f)-BNC(f)
P/N 1250-0080
BNC(f)-BNC(f)
P/N 1250-0080
Wire and
soldering
Wire and
soldering
DUT
North Hills
balun
North Hills
balun
Figure 7. Measurement configuration (3)
3
935955129.119.png 935955129.120.png 935955129.121.png 935955129.122.png 935955129.123.png 935955129.124.png 935955129.125.png 935955129.127.png 935955129.128.png 935955129.129.png 935955129.130.png 935955129.131.png 935955129.132.png 935955129.133.png 935955129.134.png 935955129.135.png 935955129.136.png 935955129.138.png 935955129.139.png 935955129.140.png 935955129.141.png 935955129.142.png 935955129.143.png 935955129.144.png 935955129.145.png 935955129.146.png 935955129.147.png 935955129.002.png 935955129.003.png 935955129.004.png 935955129.005.png 935955129.006.png 935955129.007.png 935955129.008.png 935955129.009.png 935955129.010.png 935955129.011.png 935955129.013.png 935955129.014.png 935955129.015.png 935955129.016.png 935955129.017.png 935955129.018.png 935955129.019.png 935955129.020.png 935955129.021.png 935955129.022.png 935955129.024.png 935955129.025.png 935955129.026.png 935955129.027.png 935955129.028.png 935955129.029.png 935955129.030.png 935955129.031.png 935955129.032.png 935955129.033.png
Appendix: Agilent Balun
Transformer Information
The Agilent balun transformers are
excellent interfaces for measuring
balanced components or circuits.
They can be used with unbalanced
system measurement instruments
such as impedance analyzers, LCR
meters, and/or network analyzers
(that are unbalanced system measure-
ment instruments). Agilent offers the
following balun transformers.
Figure 8. Agilent 16314-60011 and 16315-60011
Agilent 16314-60011
Agilent 16315-60011, 16316A, 16317A
The 16314-60011 can be directly
connected to a 4-terminal-pair imped-
ance analyzer or an LCR meter. High-
accuracy impedance measurements of
balanced devices can be made by
using the OPEN/SHORT/LOAD char-
acteristic. This unique feature of the
Agilent impedance analyzer and LCR
meters is performed at the binding
posts by using the furnished shorting
plate and a 50
• Reflection and transmission
measurements of balanced devices
and circuits can be measured with a
network analyzer.
• With the furnished load resistor
(50
) and short
plate, calibration can be performed
for high-accuracy measurements.
• Depending on the impedance of the
device, balanced impedance of
50
, 100
, 600
load resistor.
, 100
, or 600
can be
selected.
• Wide frequency range (100 Hz to
10 MHz (Agilent 16317A covers up
to 3 MHz))
• Recommended instruments:
• 4395A 500 MHz network/
spectrum/impedance analyzer
• 4396B 1.8 GHz network/
spectrum/impedance analyzer
• 4294A precision impedance
analyzer
• Wide frequency range
(100 Hz to 10 MHz) is covered.
• Recommended instruments:
• 4294A precision impedance
analyzer
• 4284A precision LCR meter
• 4285A precision LCR meter
• E4980A precision LCR meter
4
935955129.035.png 935955129.036.png
Specifications
Specifications describe the instrument’s war-
ranted performance over the temperature
range of 0 to 50 °C (except where noted) and
after 30-minute warm-up time.
Table 2. Agilent balun transformers specifications
Item
16314-60011
16315-60011
16316A
16317A
Binding posts
2 signal terminals and 1 ground terminal (signal terminal spacing: 14.0 mm)
Terminal
configuration &
nominal
characteristic
impedance
Balanced port
50
50
100
600
Unbalanced
port
4 BNC connectors
1 BNC connector
50
Size (mm)
Weight (g)
Operating temperature
89(W) x 56(H) x 133 (D)
400
89(W) x 55(H) x 121 (D)
350
0 to 55
°
C
Operating humidity
Non-operation temperature
Non-operating humidity
95% RH (@ 40
°
C)
-40 to +70 ° C
90% RH (@ 65
°
C)
50 load resistor
50 load resistor
100 load resistor
600 load resistor
Furnished accessories
Shorting plate
Shorting plate
Shorting plate
Shorting plate
Manual
Manual
Manual
Manual
Supplemental Characteristics
Supplemental characteristics are intended
to provide information useful in applying
the instrument by giving non-warranted
performance parameters.
Table 3. Agilent Balun transformers
supplemental characteristics (at 23 ± 5 °C)
Item
16314-60011
16315-60011
16316A
16317A
Operating
frequency
100 Hz to 10 MHz
100 Hz to 3 MHz
1.0 dB (@100 kHz)
1.0 dB (relative to the insertion loss at
100 kHz)
Insertion loss
1.5 dB (relative to the insertion loss at
100 kHz)
±
Freq. response
10 dB (100 Hz Freq. < 300 Hz)
Freq. < 300 MHz)
20 dB (300 MHz Freq. 1 MHz)
10 dB (100 Hz
Freq. < 7 MHz)
17 dB (7 MHz Freq. < 10 MHz)
50 dB (100 Hz Freq. < 3 MHz)
20 dB (300 Hz
Return loss
15 dB (1 MHz < Freq.
3MHz)
Freq. < 1 MHz)
45 dB (1 MHz Freq. < 3 MHz)
50 dB (100 Hz
Common mode
loss
45 dB (3 MHz
Freq. < 5 MHz)
40 dB (5 MHz Freq. < 10 MHz)
5
935955129.037.png 935955129.038.png 935955129.039.png 935955129.040.png 935955129.041.png 935955129.042.png 935955129.043.png 935955129.044.png 935955129.046.png 935955129.047.png 935955129.048.png 935955129.049.png 935955129.050.png 935955129.051.png 935955129.052.png 935955129.053.png 935955129.054.png 935955129.055.png 935955129.057.png 935955129.058.png 935955129.059.png 935955129.060.png 935955129.061.png
Zgłoś jeśli naruszono regulamin